R&S ESR Series User Manual page 730

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Manual operation:
Table 11-21: Return values for TRACE1 to TRACE6 parameter
The trace data consists of a list of power levels that have been measured. The number of power levels in
the list depends on the currently selected number of sweep points. The unit depends on the measurement
and on the unit you have currently set.
If you are measuring with the auto peak detector, the command returns positive peak values only.
If you are working with the "Real/Imag (I/Q)" result display in I/Q Analyzer mode, the command returns two
values for each trace point, first the real parts of the signal, then the imaginary parts (I
For SEM or Spurious Emission measurement results, the x-values should be queried as well, as they are
not equi-distant (see
Table 11-22: Return values for LIST parameter
For every measurement range you have defined (range 1...n), the command returns eight values in the fol-
lowing order.
<No>,<StartFreq>,<StopFreq>,<RBW>,<PeakFreq>,<PowerAbs>,<PowerRel>,<PowerDelta>,<Limit-
Check>,<Unused1>,<Unused2>
11.5.1.2
Measuring Channel Power
The following commands configure and perform channel power or ACLR measure-
ments.
Managing Measurement
Configuring ACLR
Defining Weighting
Selecting the Reference
Checking
Analyzing
Predefined CP/ACLR
Managing Measurement Configurations
The following commands control measurement configurations for ACLR measure-
ments.
CALCulate<n>:MARKer<m>:FUNCtion:POWer:PRESet
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:CATalog?
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:DELete
CALCulate<n>:MARKer<m>:FUNCtion:POWer:STANdard:SAVE
User Manual 1175.7068.02 ─ 12
See
See
TRACe<n>[:DATA]:X?
<No>: range number
<StartFreq>,<StopFreq>: start and stop frequency of the range
<RBW>: resolution bandwidth
<PeakFreq>: frequency of the peak in a range
<PowerAbs>: absolute power of the peak in dBm
<PowerRel>: power of the peak in relation to the channel power in dBc
<PowerDelta>: distance from the peak to the limit line in dB, positive values indicate a failed limit
check
<LimitCheck>: state of the limit check (0 = PASS, 1 = FAIL)
<Unused1>,<Unused2>: reserved (0.0)
Channels................................................................................. 729
Filters..................................................................................... 734
Limits.....................................................................................................737
Results..................................................................................................745
Standards...........................................................................748
Remote Commands in Spectrum Analyzer Mode
"List Evaluation (On/Off)"
"List Evaluation (On/Off)"
on page 856).
Configurations............................................................... 727
Channel......................................................................... 736
Remote Control
on page 264
on page 293
,...,I
1
.................................................... 728
.................................. 729
..................................... 729
........................................729
, Q
,...,Q
).
n
1
n
727

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents