Source RF Internal (On/ Off)
Switches the selected tracking generator on or off. Default setting is off. The currently
selected generator is indicated on the softkey.
Remote command:
OUTPut[:STATe]
Activates the internal tracking generator (B9).
SOURce<n>:EXTernal<generator>[:STATe]
Activates the external tracking generator 1 (B10).
Source Power
Opens an edit dialog box to quickly change the output power of the currently selected
tracking generator, alternatively to the "Tracking Generator configuration" dialog box
(see
The default output power is -20 dBm. The range is specified in the data sheet.
Remote command:
SOURce<n>:POWer[:LEVel][:IMMediate][:AMPLitude]
Source Cal
Opens a submenu to configure calibration for transmission and reflection measurement
for tracking generators. For details on the test setups see
for transmission and reflection
Calibrate Transmission ← Source Cal
Starts a sweep that records a reference trace. This trace is used to calculate the differ-
ence for the normalized values.
Remote command:
[SENSe:]CORRection:METHod
Selects the transmission method and starts the sweep to record a reference trace.
Calibrate Reflection Short ← Source Cal
Starts a sweep as a reference trace for short-circuit calibration.
If both calibrations (open circuit, short circuit) are carried out, the calibration curve is
calculated by averaging the two measurements and stored in the memory. The order of
the two calibration measurements is irrelevant.
Remote command:
[SENSe:]CORRection:METHod
Selects the reflection method.
[SENSe:]CORRection:COLLect[:ACQuire]
Starts the sweep for short-circuit calibration.
Calibrate Reflection Open ← Source Cal
Starts a sweep as a reference trace for the open-circuit calibration.
User Manual 1175.7068.02 ─ 12
Sweep.............................................................................................................. 480
Power Sweep (On
Power Sweep
Start....................................................................................... 480
Power Sweep
Stop....................................................................................... 480
Config............................................................................................................. 481
on page 901
"Source Config"
on page 481 softkey).
Softkeys of the Tracking Generator Menu
/Off).................................................................................480
measurement",
on page 493.
on page 902
on page 902
on page 901
Tracking Generator
on page 912
on page 905
Chapter 8.4.2, "Calibrating
477