Example:
Manual operation:
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO <State>
This command couples or decouples the resolution bandwidth to the selected fre-
quency.
The resolution bandwidth is coupled to the frequency only if you have selected the
Quasipeak, CISPR Average or CISPR RMS detector.
For more information see
Parameters:
<State>
Example:
Manual operation:
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE <FilterType>
This command selects the type of resolution filter.
For detailed information on filters see
Type",
on page 367.
When changing the filter type, the next larger filter bandwidth is selected if the same
filter bandwidth is not available for the new filter type.
5 Pole filters are not available when using the sweep type "FFT".
Parameters:
<FilterType>
User Manual 1175.7068.02 ─ 12
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
See
"Res BW Manual"
See
"Res BW CISPR / CISPR (6 dB)"
See
"Res BW Mil Std/ MIL Std (6 dB)"
See
"Res BW"
See
"Res BW Manual"
Chapter 5.2.1, "Measurement
ON | OFF
*RST:
BAND:AUTO OFF
Decouples the measurement bandwidth from the frequency
range.
See
"CISPR RBW Uncoupled"
on page 366 and
Chapter 6.2.6.4, "List of Available RRC and Channel
NORMal
Gaussian filters
CFILter
Channel filters
RRC
RRC filters
P5
5 Pole filters
Remote Commands in Receiver Mode
on page 184
on page 316
on page 361
Bandwidth",
ON
on page 184
Chapter 6.2.6.3, "Selecting the Appropriate Filter
Remote Control
on page 301
on page 301
on page 166.
Filters",
671