Remote command:
DISP:TRAC OFF, see
Hold/Cont ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Switches the reset of the traces in Min Hold, Max Hold and Average mode after some
specific parameter changes have been made on and off. The default setting is off.
Normally, the measurement is started anew after parameter changes, before the mea-
surement results are evaluated (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
Remote command:
DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous
Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Opens a submenu to select the detector manually, or activate automatic selection.
Note: When measuring spurious emissions, using this softkey automatically opens the
Sweep List dialog, see
If a detector was selected manually, the "MAN" indicator is highlighted.
If "AUTO" is selected, the detector is defined automatically, depending on the selected
trace mode:
Trace mode
Clear Write
Max Hold
Min Hold
Average
View
Blank
Auto Select ← Detector ← Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the best detector for the selected trace and filter mode. This is the default set-
ting.
Trace mode
Clear/Write
Average
Max Hold
Min Hold
Remote command:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]:AUTO
User Manual 1175.7068.02 ─ 12
DISPlay[:WINDow<n>]:TRACe<t>[:STATe]
"Sweep List dialog box"
Detector
Auto Peak
Positive Peak
Negative Peak
Sample Peak
Detector
Auto Peak
Sample
Max Peak
Min Peak
Spectrum Measurements
on page 849
on page 289.
on page 853
Analysis
on page 849
391