Example:
Manual operation:
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MAXimum <Level>
This command sets the upper limit level for one power class. The unit is dBm. The limit
always ends at + 200 dBm, i.e. the upper limit of the last power class can not be set. If
more than one power class is in use, the upper limit must equal the lower limit of the
next power class.
Suffix:
<n>
<k>
<Class>
Parameters:
<Level>
Example:
Manual operation:
CALCulate<n>:LIMit<k>:ESPectrum:PCLass<Class>:MINimum <Level>
This command sets the minimum lower level limit for one power class. The unit is dBm.
The limit always start at – 200 dBm, i.e. the first lower limit can not be set. If more than
one power class is in use, the lower limit must equal the upper limit of the previous
power class.
Suffix:
<n>
<k>
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
ABSolute
Evaluates only limit lines with absolute power values
RELative
Evaluates only limit lines with relative power values
AND
Evaluates limit lines with relative and absolute power values. A
negative result is returned if both limits fail.
OR
Evaluates limit lines with relative and absolute power values. A
negative result is returned if at least one limit failed.
*RST:
REL
CALC:LIM:ESP:PCL:LIM ABS
See
"Used Power Classes"
.
irrelevant
irrelevant
1...4
the power class to be evaluated
<numeric value>
*RST:
+200
CALC:LIM:ESP:PCL1:MAX -40 dBm
Sets the maximum power value of the first power class to -40
dBm.
See
"PMin/PMax"
on page 266
.
irrelevant
irrelevant
Remote Control
on page 266
775