R&S ESR Series User Manual page 817

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
The function is only available in the REMOTE operation. It is deactivated on switching
back to LOCAL.
Parameters:
<Frequency>
<RBW>
<MeasTime>
<TriggerSource>
<TriggerLevel>
<TriggerOffset>
<Detector>
<#OfPulses>
Return values:
<PowerLevel>
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
Defines the pulse frequency.
Range:
0 Hz to fmax
Defines the resolution bandwidth.
Range:
10 Hz to 10 MHz in steps of 1, 3, 10
Defines the measurement time.
Range:
1 µs to 30 s
Selects a trigger source.
EXTernal
The trigger signal is fed from the "Ext. Trigger/Gate" input on the
rear of the unit.
VIDeo
The internal video signal is used as trigger signal.
Defines a trigger level.
The trigger level is available for the video trigger. In that case,
the level is a percentage of the diagram height.
In case of an external trigger, the R&S ESR uses a fix TTL level.
Range:
0 PCT to 100 PCT
Defines the trigger delay. The trigger delay is the time between
the detection of the trigger signal and the start of the measure-
ment.
Range:
see data sheet
Selects the detector and therefore the way the measurement is
evaluated.
MEAN
Calculates the RMS pulse power.
PEAK
Calculates the peak pulse power.
Defines the number of pulses included in the measurement.
Range:
1 to 32001
Pulse power level.
The result is a list of floating point values separated by commas.
The unit is dBm.