R&S
ESR
Suffix:
<n>
Example:
Manual operation:
[SENSe:]SWEep:COUNt <NumberSweeps>
This command defines the number of sweeps started with single sweep, which are
used for calculating the average or maximum value. If the values 0 or 1 are set, one
sweep is performed.
Parameters:
<NumberSweeps>
Example:
Manual operation:
[SENSe:]SWEep:POINts <NumberPoints>
This command defines the number of measurement points to be collected during one
sweep.
Note: For Spurious Emissions measurements the maximum number of sweep points in
all ranges is limited to 100001.
Parameters:
<NumberPoints>
Example:
Manual operation:
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
.
irrelevant
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Setting the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
INIT:CONM;*WAI
Continues the measurement (next 20 sequences) and waits for
the end.
See
"Continue Single Sweep"
0 to 32767
*RST:
0 (GSM: 200, PHN:1)
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
See
"Sweep Count"
on page 372
Range:
101 to 32001
*RST:
691
SWE:POIN 150
See
"Sweep Points"
on page 302
See
"Sweep Points"
on page 372
Remote Control
on page 370
834