The changes of the sweep list are only kept until you load another parameter set (by
pressing PRESET or by loading an XML file). If you want a parameter set to be availa-
ble permanently, create an XML file for this configuration (for details refer to
Description of Spectrum Emission Mask XML Files"
If you load one of the provided XML files ("Load Standard" softkey, see
dard"
selected standard. For further details refer also to
trum Emission Mask Measurement"
Note: If you edit the sweep list, always follow the rules and consider the limitations
described in
Range Start / Range Stop ← Sweep List dialog box ← Sweep List ← Spectrum
Emission Mask
Sets the start frequency/stop frequency of the selected range. Follow the rules descri-
bed in
In order to change the start/stop frequency of the first/last range, select the appropriate
span with the SPAN key. If you set a span that is smaller than the overall span of the
ranges, the measurement includes only the ranges that lie within the defined span and
have a minimum span of 20 Hz. The first and last ranges are adapted to the given
span as long as the minimum span of 20 Hz is not violated.
Frequency values for each range have to be defined relative to the center frequency.
The reference range has to be centered on the center frequency. The minimum span of
the reference range is given by the current TX Bandwidth.
Remote command:
[SENSe:]ESPectrum:RANGe<range>[:FREQuency]:STARt
[SENSe:]ESPectrum:RANGe<range>[:FREQuency]:STOP
Fast SEM ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Activates "Fast SEM" mode for all ranges in the sweep list. For details see
trum Emission Mask Measurements"
Note: If "Fast SEM" mode is deactivated while
rical Setup" mode is automatically also deactivated.
If "Fast SEM" mode is activated while "Symmetrical Setup" mode is on, not all range
settings can be set automatically.
Remote command:
[SENSe:]ESPectrum:HighSPeed
Filter Type ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Sets the filter type for this range. For details on filters see also
"Selecting the Appropriate Filter
Remote command:
[SENSe:]ESPectrum:RANGe<range>:FILTer:TYPE
RBW ← Sweep List dialog box ← Sweep List ← Spectrum Emission Mask
Sets the RBW value for this range.
Remote command:
[SENSe:]ESPectrum:RANGe<range>:BANDwidth[:RESolution]
User Manual 1175.7068.02 ─ 12
on page 267), the sweep list contains ranges and parameters according to the
"Ranges and Range Settings"
"Ranges and Range Settings"
on page 275).
"Provided XML Files for the Spec-
on page 272.
on page 269.
on page 269.
on page 281.
Symmetric Setup
on page 762
Type",
on page 366.
Spectrum Measurements
Measurements
"Format
"Load Stan-
on page 765
on page 765
"Fast Spec-
mode is on, "Symmet-
Chapter 6.2.6.3,
on page 764
on page 763
259

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents