R&S ESR Series User Manual page 893

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
TRACe<n>:IQ:RLENgth <NoOfSamples>
This command sets the record length for the acquired I/Q data. Increasing the record
length automatically also increases the measurement time.
Note: Alternatively, you can define the measurement time using the SENS:SWE:TIME
command.
Suffix:
<n>
Parameters:
<NoOfSamples>
Example:
Manual operation:
TRACe<n>:IQ:SET NORM,<Placeholder>, <SampleRate>, <TriggerMode>,
This command defines the settings of the R&S ESR hardware for the measurement of
I/Q data. This allows setting the sample rate, trigger conditions and the record length.
If this command is omitted, the current instrument settings are used for the correspond-
ing parameters.
This command switches to IQ mode automatically (see also
on page 898).
The trigger level can be set using the
Note: When using the default settings with the
following minimum buffer sizes for the response data are recommended: ASCII format
10 kBytes, Binary format 2 kBytes.
Tip: To determine the trigger point in the sample (for "External" or "IF Power" trigger
mode), use the
Parameters:
NORM
<Placeholder>
<TriggerMode>
User Manual 1175.7068.02 ─ 12
.
irrelevant
Number of measurement values to record.
Range:
*RST:
TRAC:IQ:RLEN 256
See
"Record Length"
<TriggerSlope>, <PretriggerSamp>, <NumberSamples>
TRACe<n>:IQ:TPISample?
is not evaluated, but must be inserted
numeric value
is not evaluated, but must be inserted
IMMediate | EXTernal | IFPower | BBPower | PSE
Selection of the trigger source used for the measurement.
For IMM mode, gating is automatically deactivated.
*RST:
Remote Commands in I/Q Analyzer Mode
1 ... 209715200(= 200*1024*1024);
691
on page 460
TRIGger<n>[:SEQuence]:SOURce
TRACe<n>:IQ:DATA
on page 899 command.
IMM
Remote Control
TRACe<n>:IQ[:STATe]
command.
command, the