R&S ESR Series User Manual page 673

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
INPut:IMPedance <Impedance>
This command selects the nominal input impedance.
75 Ω should be selected if the 50 Ω input impedance is transformed to a higher impe-
dance using a 75 Ω adapter of the RAZ type (= 25 Ω in series to the input impedance
of the instrument). The correction value in this case is 1.76 dB = 10 log (75Ω/50Ω).
Parameters:
<Impedance>
Example:
Manual operation:
11.4.4 Selecting the Bandwidth
[SENSe:]BANDwidth:IF
[SENSe:]BANDwidth|BWIDth[:RESolution]
[SENSe:]BANDwidth|BWIDth[:RESolution]:AUTO
[SENSe:]BANDwidth|BWIDth[:RESolution]:TYPE
[SENSe:]BANDwidth:IF <Bandwidth>
This command defines the resolution bandwidth for IF analysis.
Parameters:
<Bandwidth>
Example:
Manual operation:
[SENSe:]BANDwidth|BWIDth[:RESolution] <Bandwidth>
This command defines the resolution bandwidth.
The available bandwidths depend on the selected filter type. For more information see
Chapter 5.2.1, "Measurement
A change of the resolution bandwidth automatically turns the coupling to the span off.
Parameters:
<Bandwidth>
User Manual 1175.7068.02 ─ 12
50 | 75
*RST:
50 Ω
INP:IMP 75
See
"Input 50 Ω/75 Ω "
...................................................................................................670
....................................................................... 670
Resolution filter bandwidth with a 6 dB bandwidth in the range
from 10 Hz to 100 kHz in steps of 1-3-10-....
The available bandwidths depend on the current frequency
span.
*RST:
3 kHz
BAND:IF 10KHZ
Defines a resolution bandwidth of 10 kHz.
See
"IF Analysis RBW"
Bandwidth",
on page 166.
refer to data sheet
*RST:
(AUTO is set to ON)
Remote Commands in Receiver Mode
on page 184
..............................................................671
.............................................................. 671
on page 185
Remote Control
670

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents