R&S ESR Series User Manual page 769

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
[SENSe:]ESPectrum:RANGe<range>:INPut:ATTenuation <Value>
This command sets the attenuation for the specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
Parameters:
<Value>
Example:
Manual operation:
[SENSe:]ESPectrum:RANGe<range>:INPut:ATTenuation:AUTO <State>
This command activates or deactivates the automatic RF attenuation setting for the
specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
Parameters:
<State>
Example:
Manual operation:
[SENSe:]ESPectrum:RANGe<range>:INPut:GAIN:STATe <State>
This command switches the preamplifier on or off for the specified range.
Note that this parameter can not be set for all ranges if "Fast SEM" mode is activated
(see
Suffix:
<range>
Parameters:
<State>
User Manual 1175.7068.02 ─ 12
[SENSe:]ESPectrum:HighSPeed
.
1...20
range
Refer to the data sheet.
*RST:
ESP:RANG3:INP:ATT 10
Sets the attenuation of range 3 to 10 dB.
See
"RF Attenuator"
[SENSe:]ESPectrum:HighSPeed
.
1...20
range
ON | OFF
*RST:
ESP:RANG2:INP:ATT:AUTO OFF
Deactivates the RF attenuation auto mode for range 2.
See
"RF Att. Mode"
[SENSe:]ESPectrum:HighSPeed
.
1...20
range
ON | OFF
*RST:
Remote Commands in Spectrum Analyzer Mode
on page 762).
0 dB
on page 260
on page 762).
ON
on page 260
on page 762).
OFF
Remote Control