CALCulate<n>:STATistics:NSAMples <NoMeasPoints>
This command sets the number of measurement points to be acquired for the statisti-
cal measurement functions.
Suffix:
<n>
Parameters:
<NoMeasPoints>
Example:
Manual operation:
Retrieving Results
The following commands are required to retrieve the measurement results.
Commands useful to retrieve results described elsewhere:
CALCulate<n>:MARKer<m>:X
TRACe<n>:DATA
CALCulate<n>:STATistics:CCDF:X<Trace>
CALCulate<n>:STATistics:RESult<Trace>
CALCulate<n>:STATistics:CCDF:X<Trace> <Probability>
This command queries the results of the complementary cumulative distribution func-
tion (CCDF) for the specified probability (corresponding to the "Percent Marker" in the
display, see
Suffix:
<n>
<Trace>
Parameters:
<Probability>
Example:
Manual operation:
User Manual 1175.7068.02 ─ 12
.
irrelevant
100 to 1E9
*RST:
CALC:STAT:NSAM 500
Sets the number of measurement points to be acquired to 500.
See
"# of Samples"
on page 726
"Percent Marker"
on page 322).The suffix of X defines the trace number.
.
irrelevant
1...6
trace
P0_01: Level value for 0.01 % probability
P0_1: Level value for 0.1 % probability
P1: Level value for 1 % probability
P10: Level value for 10 % probability
CALC:STAT:CCDF:X1? P10
Returns the level values that are over 10 % above the mean
value.
See
"CCDF"
Remote Commands in Spectrum Analyzer Mode
100000
on page 317
on page 695
...................................................................... 793
........................................................................ 794
on page 321
Remote Control