"Next Min"
"Next Mode"
"Search Limits"
"Peak Excursion"
Spectrogram:
Functions to position markers exclusive in Spectrum mode
Center =Mkr Freq (span >
Ref Lvl =Mkr
Auto Max Peak/Auto Min
Exclude
Center =Mkr Freq (span > 0)
Sets the center frequency to the current marker or delta marker frequency. A signal
can thus be set to as center frequency, for example to examine it in detail with a
smaller span.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:CENTer
Ref Lvl =Mkr Lvl
Sets the reference level to the current marker level.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
Auto Max Peak/Auto Min Peak
Adds an automatic peak search action for marker 1 at the end of each particular
sweep. This function may be used during adjustments of a device under test to keep
track of the current peak marker position and level.
The current marker search limit settings
taken into account.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:AUTO
CALCulate<n>:MARKer<m>:MINimum:AUTO
Exclude LO
Switches the frequency range limit for the marker search functions on or off.
"ON"
"OFF"
Remote command:
CALCulate<n>:MARKer<m>:LOEXclude
User Manual 1175.7068.02 ─ 12
on page 205
on page 205
on page 206
on page 207
Chapter 5.4.2.3, "Markers in
0).......................................................................................412
Lvl.......................................................................................................... 412
Peak................................................................................... 412
LO..................................................................................................................412
The minimum frequency included in the peak search range is ≥ 5 ×
resolution bandwidth (RBW).
Due to the interference by the first local oscillator to the first inter-
mediate frequency at the input mixer, the LO is represented as a sig-
nal at 0 Hz. To avoid the peak marker jumping to the LO signal at 0
Hz, this frequency is excluded from the peak search.
No restriction to the search range. The frequency 0 Hz is included in
the marker search functions.
Spectrum Measurements
Spectrograms",
on page 688
on page 862
(Left
Limit,
Right
Limit,
on page 863
on page 863
on page 863
Analysis
on page 207
Threshold
softkeys) are
412