R&S ESR Series User Manual page 878

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Suffix:
<n>
<m>
Parameters:
<Duration>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect <DemodMode>
This command selects the demodulation mode for the audio demodulator.
Suffix:
<n>
<m>
Parameters:
<DemodMode>
Example:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe] <State>
This command turns the audio demodulator on and off when the measurement rea-
ches a marker position.
In the frequency domain, the hold time can be defined at the corresponding marker
position with CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff.
In the time domain continuous demodulation is always on.
Suffix:
<n>
<m>
Parameters:
<State>
Example:
Manual operation:
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
.
Selects the measurement window.
Selects the marker.
Range:
10 ms to 1000 s
*RST:
Marker demodulation = OFF
CALC:MARK:FUNC:DEM:HOLD 3s
See
"Mkr Stop Time"
on page 420
.
Selects the measurement window.
irrelevant
AM | FM
*RST:
AM
CALC:MARK:FUNC:DEM:SEL FM
See
"AM"
on page 420
See
"FM"
on page 420
.
Selects the measurement window.
Selects the marker.
ON | OFF
*RST:
OFF
CALC:MARK3:FUNC:DEM ON
Switches on the demodulation for marker 3.
See
"Mkr Demod On/Off"
Remote Control
on page 420
875

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents