2. Set the center frequency to 128 MHz and the frequency span to 100 MHz:
a) Press the FREQ key and enter "128 MHz".
b) Press the SPAN key and enter "100 MHz".
3. To suppress noise spikes the trace can be averaged.
a) Press the TRACE key.
b) Press the "Trace Wizard" softkey.
User Manual 1175.7068.02 ─ 12
Figure 6-25: Sine wave signal with low S/N ratio. The signal is measured with the auto peak
detector and is completely hidden in the intrinsic noise of the R&S ESR.
The Trace Wizard dialog box opens.
Spectrum Measurements
Advanced Measurement Examples
438