R&S ESR Series User Manual page 485

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Figure 8-1: TTL connection for an SMU generator
In order to enhance measurement accuracy, a common reference frequency should be
used for both the R&S ESR and the generator. If no independent 10 MHz reference
frequency is available, it is recommended that you connect the reference output of the
generator with the reference input of the R&S ESR and that you enable usage of the
external reference on the R&S ESR via "SETUP" >"Reference EXT".
8.2.1 Overview of Generators Supported by the R&S ESR
The R&S SMA and R&S SMU require the following firmware versions:
R&S SMA: V2.10.x or higher
R&S SMU: V1.10 or higher
Generator
SMA01A
SMBV100A3
SMBV100A6
SMC100A1
1) Requires the option SMR-B11 to be fitted.
2) Maximum/minimum power depends on presence of Option SMR-B15/-B17 and set frequency range. For
more details see SMR data sheet.
3) Maximum/minimum power depends on presence of Option SMP-B15/-B17 and set frequency range. For
more details see SMP data sheet.
4) Requires firmware version V3.20.200 or higher on the signal generator
User Manual 1175.7068.02 ─ 12
Interface Type
Generator Min
Freq
TTL
9 kHz
TTL
9 kHz
TTL
9 kHz
TTL
9 kHz
Tracking Generator
Connecting External Tracking Generators
Generator Max
Generator Min
Freq
Power
dBm
3.0 GHz
-145
3.2 GHz
-145
6.0 GHz
-145
1.1 GHz
-120
Generator Max
Power
dBm
+30
+30
+30
+30
482

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents