R&S ESR Series User Manual page 199

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Final Measurement Time
Defines the measurement time for the final measurement.
The measurement time is the time during which the R&S ESR measures the input sig-
nal and forms a measurement result weighted by the selected detector. The measure-
ment time does not include settling times of the synthesizer and the IF filter. The
R&S ESR automatically waits until transients are over.
Interactive Mode
Turns interactive final measurements on and off.
Remote command:
[SENSe:]FMEasurement:AUTO
Trace 1 to 6
Selects the characteristics of each trace.
In addition to the trace mode, you can select a detector for the scan and the final mea-
surement. For more information see
Chapter 5.1.3,
Chapter 5.1.5, "Final
Chapter 5.2.3, "Trace
Chapter 5.2.2,
Note: When you select detector type "None" for the final measurement, the R&S ESR
ignores peaks found on the corresponding trace during the final measurement.
Remote command:
Trace mode:
DISPlay[:WINDow<n>]:TRACe<t>:MODE
Scan detector:
[SENSe:][WINDow:]DETector<trace>[:FUNCtion]
Final measurement detector:
[SENSe:]DETector<t>:FMEasurement
5.3.7.6
LISN Settings
The "LISN Settings" tab in the "Test Automation" dialog box contains functionality to
control line impedance networks.
User Manual 1175.7068.02 ─ 12
on page 643
"Scans",
on page 144
Measurement",
on page 150
Modes",
on page 171
"Detectors",
on page 167
Common Measurement Settings
on page 683
on page 683
on page 642
Receiver Mode