R&S ESR Series User Manual page 721

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Suffix:
<n>
<k>
Example:
Usage:
CALCulate<n>:LIMit<k>:FAIL?
This command queries the result of a limit check.
Note that for SEM measurements, the limit line suffix <k> is irrelevant, as only one spe-
cific SEM limit line is checked for the currently relevant power class.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweeps.
Suffix:
<n>
<k>
Return values:
<Result>
Example:
Usage:
Manual operation:
CALCulate<n>:LIMit<k>:STATe <State>
This command turns the limit check on and off.
To query the limit check result, use CALCulate<n>:LIMit<k>:FAIL?.
Suffix:
<n>
<k>
Parameters:
<State>
User Manual 1175.7068.02 ─ 12
.
irrelevant
irrelevant
CALC:LIM:CLE
Deletes the result of the limit check.
Event
.
irrelevant
limit line
0
PASS
1
FAIL
INIT;*WAI
Starts a new sweep and waits for its end.
CALC:LIM3:FAIL?
Queries the result of the check for limit line 3.
Query only
See
"Limit Check 1-4"
See
"Limit Check 1-4"
.
irrelevant
Selects the limit line.
ON | OFF
*RST:
OFF
Remote Commands in Receiver Mode
on page 261