Return values:
<Condition>
Example:
Usage:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:LIMit<k>:DELTa?
This command queries the vertical distance from the marker position to the limit line.
The unit is dB.
If the marker has been assigned to a different trace than the limit line, the command
returns -200.
Suffix:
<n>
<m>
<k>
Return values:
<Amplitude>
Example:
Usage:
Manual operation:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:RESult? <Result>
This command queries the result of the EMI measurement at the marker position.
Suffix:
<n>
User Manual 1175.7068.02 ─ 12
1...16
marker
1...8
limit line
0
The marker has passed the limit check.
1
The marker is inside the margins of a limit line.
2
The marker has failed the limit check.
CALC:MARK1:FUNC:FME:LIM2:COND?
Queries the condition of marker 1 in relation to limit line 2.
Query only
See
"Marker Table"
<Amplitude>
.
irrelevant
1...16
marker
1...8
limit line
Vertical distance to the limit line in dB.
CALC:MARK3:FUNC:FME:LIM2:DELT?
Queries the distance of marker 3 to the second limit line.
Query only
See
"Marker Table"
.
irrelevant
Remote Commands in Spectrum Analyzer Mode
on page 305
on page 305
Remote Control
783