Figure 6-8: CCDF of white noise
A red line indicates the ideal Gaussian distribution for the measured amplitude range
(white noise).
The displayed amplitude range is indicated as "Mean Pwr + <x dB>"
In addition to the histogram, a result table is displayed containing the following informa-
tion:
Number of samples used for calculation
For each displayed trace:
Mean
Peak
Crest
0,01 %
0,1 %
User Manual 1175.7068.02 ─ 12
Mean power
Peak power
Crest factor (peak power – mean power)
Level values over 0,01 % above mean power
Level values over 0,1 % above mean power
Spectrum Measurements