R&S ESR Series User Manual page 797

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
CALCulate<n>:STATistics:RESult<Trace> <ResultType>
This command reads out the results of statistical measurements of a recorded trace.
Suffix:
<n>
<Trace>
Parameters:
<ResultType>
Example:
Using Gate Ranges for Statistical Measurements
The following commands control gated statistical measurements.
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent
[SENSe:]SWEep:EGATe:TRACe<k>: PERiod
[SENSe:]SWEep:EGATe:TRACe<k>:STARt<range>
[SENSe:]SWEep:EGATe:TRACe<k>[:STATe<range>]
[SENSe:]SWEep:EGATe:TRACe<k>: STOP<range>
[SENSe:]SWEep:EGATe:TRACe<k>:COMMent <Comment>
Defines a comment for one of the traces for gated triggering.
Suffix:
<k>
Parameters:
<Comment>
Example:
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
.
irrelevant
1...6
trace
MEAN | PEAK | CFACtor | ALL
MEAN
Average (=RMS) power in dBm measured during the measure-
ment time.
PEAK
Peak power in dBm measured during the measurement time.
CFACtor
Determined CREST factor (= ratio of peak power to average
power) in dB.
ALL
Results of all three measurements mentioned before, separated
by commas: <mean power>,<peak power>,<crest factor>
The required result is selected via the following parameters:
CALC:STAT:RES2? ALL
Reads out the three measurement results of trace 2. Example of
answer string: 5.56,19.25,13.69 i.e. mean power: 5.56 dBm,
peak power 19.25 dBm, CREST factor 13.69 dB
................................................................ 794
................................................................... 795
.
1...6
trace
SWE:EGAT:TRAC1:COMM "SlotA"
Remote Control
..........................................................795
........................................................ 795
......................................................... 796
794

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents