R&S ESR Series User Manual page 104

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR
Figure 3-11: Measurement of two equally-leveled RF sinusoidal signals with the resolution band-
Matching generator and R&S ESR frequencies
The level drop is located exactly in the center of the screen only if the generator
frequencies match the frequency display of the R&S ESR exactly. To achieve exact
matching, the frequencies of the generators and the R&S ESR must be synchron-
ized.
5. Set the resolution bandwidth to 100 kHz. To do so, in the bandwidth menu, press
the "Res BW Manual" softkey and enter 100 kHz.
It is no longer possible to clearly distinguish the two generator signals.
User Manual 1175.7068.02 ─ 12
width which corresponds to the frequency spacing of the signals
Quick Start
Basic Measurement Examples
101