Receiver Mode; Measurements And Result Displays; Bargraph Measurement - R&S ESR Series User Manual

Emi test receiver
Table of Contents

Advertisement

®
R&S
ESR

5 Receiver Mode

In receiver mode, the R&S ESR measures the level at the set frequency with a
selected bandwidth and measurement time. Signal weighting is by means of detectors.
A frequency scan can be performed after setting the start and stop frequency and the
step width. Scan subranges are defined in a table.
Functions for data reduction and the control of line impedance simulating network are
available.
Preselection is always switched on in receiver mode.

5.1 Measurements and Result Displays

The R&S ESR provides several types of measurements.
The bargraph result display shows the signal level on a single frequency.
Fast prescans together with several data reduction methods are used to reduce the
amount of data in the final measurements. The final measurement is then per-
formed only on frequencies that have a high interferer level.
The final measurement may be either automatic or in interactive mode. Automatic
control functions for line impedance stabilization networks (LISN) are available.
Time domain scans (option R&S ESR-K53) reduce the required overall measure-
ment times enormously by using fast fourier transform (FFT) of frequency sections.
Time domain scan may even eliminate the need for preliminary measurements.
Fixed frequency mode for click analysis and IF spectrum analysis mode for manual
tuning are available.
The spectrogramm result display provides an additional data evaluation method by
showing the signals over time.
Scans.................................................................................................................... 144
Spectrogram..........................................................................................................151
Measurement

5.1.1 Bargraph Measurement

The bargraph result display shows the signal level on a single frequency. It is a basic
result display that indicates the signal level numerically and graphically.
The length of the bar represents the signal level at the current receiver frequency eval-
uated with the currently selected detector.
User Manual 1175.7068.02 ─ 12
Measurement........................................................................................ 142
Analysis............................................................................................ 143
Reduction...............................................................................148
Measurement............................................................................................... 150
Control............................................................................................156
Receiver Mode
Measurements and Result Displays
142