DISPlay[:WINDow<n>]:TRACe<t>:MODE:HCONtinuous <State>
This command defines whether traces in Min Hold, Max Hold and Average mode (see
DISPlay[:WINDow<n>]:TRACe<t>:MODE
is changed.
Normally, the measurement is started anew after parameter changes, before the mea-
surement results are evaluated (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
Suffix:
<n>
<t>
Parameters:
<State>
Example:
Manual operation:
DISPlay[:WINDow<n>]:TRACe<t>[:STATe] <State>
This command switches on or off the display of the corresponding trace. The other
measurements are not aborted but continue running in the background.
Suffix:
<n>
<t>
User Manual 1175.7068.02 ─ 12
Remote Commands in Spectrum Analyzer Mode
ON | OFF
*RST:
OFF
CALC:MATH:STAT ON
Switches on the trace mathematics.
See
"Trace Math"
on page 394
See
"Trace Math Off"
on page 395
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
trace
ON
The automatic reset is switched off.
OFF
After certain parameter changes the traces are reset.
*RST:
OFF
DISP:WIND:TRAC3:MODE:HCON ON
Switches off the reset function.
See
"Hold/Cont"
on page 391
.
window; For applications that do not have more than 1 measure-
ment window, the suffix <n> is irrelevant.
trace
on page 683) are reset after a parameter