Manual operation:
INITiate<n>:CONTinuous <State>
This command turns single measurements on and off.
The measurement starts immediately.
Parameters:
<State>
Example:
Manual operation:
INITiate<n>[:IMMediate]
The command initiates a new measurement.
In case of a single measurement, the R&S ESR stops measuring when it has reached
the end frequency. When you start a continuous measurement, it stops only if you
abort it deliberately.
If you are using trace modes MAXHold, MINHold and AVERage, previous results are
reset when you restart the measurement.
In single sweep mode, you can synchronize to the end of the measurement with *OPC,
*OPC? or *WAI. In continuous sweep mode, synchronization to the end of the mea-
surement is not possible. Thus, it is not recommended that you use continuous sweep
mode in remote control, as results like trace data or markers are only valid after a sin-
gle sweep end synchronization.
Example:
Usage:
Manual operation:
User Manual 1175.7068.02 ─ 12
See
"Continue at Hold"
ON
Continuous measurements.
OFF
Single measurements.
INIT2:CONT OFF
Turns on a single measurement.
See
"Continuous Bargraph / Single Bargraph"
See
"Run Continuous / Run Single"
See
"Continuous Sweep"
See
"Single Sweep"
on page 370
INIT2:CONT OFF
Selects single measurement mode.
SWE:COUN 20
Forms an average over 20 measurements.
INIT2;*WAI
Starts the measurement and waits for the end of the complete
measurement.
Event
See
"Continue at Rec Frequency"
See
"Run Continuous / Run Single"
Remote Commands in Receiver Mode
on page 159
on page 185
on page 370
on page 159
on page 185
Remote Control
on page 158