Download Print this page

Tektronix AWG2021 User Manual page 503

Arbitrary waveform generator
Hide thumbs Also See for AWG2021:

Advertisement

Appendix C: Performance Verification
Floating Point Processor Check (Option 09)
C 50
This procedure checks the AWG2021 floating point processor. This check
requires that the AWG2021 has Option 09 installed.
Equipment Required
Prerequisites
1. Check that floating point processor test in internal diagnostics passes:
a. Run the AWG2021 internal diagnostics: Push the AWG2021 ON/STBY
switch two times so that the AWG2021 runs the power-on diagnostics.
b. Check the FPP test results: When the AWG2021 finishes the FPP test,
check that the test result is Pass.
This completes the performance tests for the AWG2021.
None.
The AWG2021 meets the prerequisites listed on page C 5.
AWG2021 User Manual

Advertisement

loading
Need help?

Need help?

Do you have a question about the AWG2021 and is the answer not in the manual?