Appendix C: Performance Verification
Check Burst Mode
C 14
Electrical
Characteristic Checked
Equipment Required
Prerequisites
1. Install test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG2021 CH1 output connector
through the coaxial cable to the CH1 vertical input connector on the
oscilloscope (see Figure C-5).
AWG2021
AWG2021
Figure C 5: Burst Mode Initial Test Hookup
b. Set the oscilloscope controls:
Vertical
. . . . . . . . . . . . . . . . . . . . . . . . .
CH1 coupling:
CH1 scale
. . . . . . . . . . . . . . . . . . . .
CH1 input impedance:
Horizontal
Sweep
. . . . . . . . . . . . . . . . . . . . . . .
Trigger
Source
. . . . . . . . . . . . . . . . . . . . . .
Coupling
. . . . . . . . . . . . . . . . . . . . .
Slope
. . . . . . . . . . . . . . . . . . . . . . .
Level
. . . . . . . . . . . . . . . . . . . . . . .
Mode
. . . . . . . . . . . . . . . . . . . . . . .
2. Set AWG2021 controls and select the waveform file:
a. Initialize AWG2021 controls: Push UTILITY!Misc!Config...!Re-
set to Factory!O.K.
b. Modify the AWG2021 default settings: Push MODE!Burst!Burst
Count and turn the general purpose knob to a burst count of 3.
Operating mode, Burst, on page B 1.
A 50 W coaxial cable and an oscilloscope.
The AWG2021 meets the prerequisites listed on page C 5.
Oscilloscope
TDS
540
CH1
. . . . . . . . . . . . . . . .
DC
0.2 V/div.
50 W
. . . . . . . . . .
500 ms/div.
CH1
DC
Positive
+100 mV
Norm