Appendix C: Performance Verification
H Set the function generator offset value to –1 V. Check that the DMM
voltage reading is in the range from –0.125 to 0.125 V (0% modulation).
6. End procedure: Keep the test connections and instrument settings for the
next check.
C 25
AWG2021 User Manual