Table B 3: Electrical Characteristics (Cont.)
Characteristics
Auxiliary Input
** TRIGGER
** Accuracy
Pulse Width
Input Swing
Maximum Input Volts
Trigger to Signal delay
Internal Clock
External Clock
Trigger Holdoff
AM
Amplitude Accuracy
Maximum Input Volts
** CLOCK
** Threshold Level
Input Swing
Pulse Width
Maximum Input Volts
Frequency Range
AM and ADD
AM (Multiply)
Output
Frequency Response
** External AM
** Sensitivity
Frequency Response
CH1
External Signal
AWG2021 User Manual
Description
(5% of Level + 0.1 V)
15 ns minimum
0.2 V
minimum
p p
10 V
when 1 MW selected
p p
5 V
when 50 W selected
RMS
100 ns maximum
100 ns maximum + 0.5 Clock + 1 Clock maximum
1 ms maximum (Excluding Autostep mode)
within 5%
5 V
0.5 V
0.8 V
minimum
p p
2 ns minimum
2 V
250 MHz maximum
When Option 02 (second channel) installed
within 5%
CH2 Signal
5 V
0 V
-5 V
DC to 30 MHz (-3 dB)
2 V
( 5%) signal causes 100% modulation.
p p
Ext Signal
1 V
0 V
-1 V
DC to 30 MHz (-3 dB)
DC to 4 MHz (-3 dB)
Appendix B: Performance Characteristics
100%
0%
-100%
100% modulation
50% modulation
0% modulation
B 9