Switching The Instrument On And Off - R&S SGT100A User Manual

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Risk of injury and instrument damage if stacking instruments
A stack of instruments may tilt over and cause injury. Furthermore, the instruments at
the bottom of the stack may be damaged due to the load imposed by the instruments
on top.
Observe the following instructions when stacking instruments:
Never stack more than three instruments with the same dimensions (width and
length). If you need to stack more than three instruments, install them in a rack.
The overall load imposed on the lowest instrument must not exceed 500 N.
All instruments should have the same dimensions (width and length).
If you need to stack smaller instruments on the top, the overall load imposed on the
lowest instrument must not exceed 250 N.
If the instruments have foldable feet, fold them in completely.
Rack mounting
The R&S SGT can be installed in a rack using a rack adapter kit (Order No. see data
sheet). The installation instructions are part of the adapter kit.
Risk of instrument damage in a rack
An insufficient airflow can cause the instrument to overheat, which may disturb the
operation and even cause damage.
Make sure that all fan openings are unobstructed, that the airflow perforations are
unimpeded, and that the minimum distance from the wall is 10 cm.

2.1.5 Switching the Instrument On and Off

The R&S SGT is automatically adapted to the AC voltage supplied. There is no need to
set the voltage manually or change fuses. The
at the rear of the unit.
User Manual 1176.8674.02 ─ 07
Preparing for Use
Putting into Operation
AC SUPPLY AND POWER SWITCH
is
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents