When to perform zeroing:
During warm up after switching on or connecting the instrument
After a substantial change of the ambient temperature
After fastening the power sensor module to an RF connector at high temperature
After several hours of operation
When low-power signals are to be measured, e.g. less than 10 dB above the lower
measurement limit.
Switch off the RF power source for zeroing, but do not disconnect it from the power
sensor. This proceeding keeps the thermal equilibrium, and the zeroing process
also compensates the noise that superimposes the measured signal (e.g. from a
broadband amplifier).