SGT100A
Remote command:
[:SOURce]:POWer:ATTenuation:SOVer[:OFFSet]
Level Range
Displays the level range in which the level is set without interruption for the "Attenuator
Mode Fixed" setting.
Remote command:
:OUTPut:AFIXed:RANGe:LOWer?
:OUTPut:AFIXed:RANGe:UPPer?
RF-Off-Mode
Determines the attenuator's state after the instrument is switched off.
In default setting, the electronic step attenuator switches to highest attenuation when
RF is off. By setting the RF-Off mode, the electronic step attenuator can be fixed to
keep the output impedance constant during RF off.
Remote command:
[:SOURce]:POWer:ATTenuation:RFOFf:MODE
7.7.6.3

Power-On Settings

► To access the power- on/EMF settings, select "SGMA-GUI > Instrument Name >
RF > Level > Power-On/EMF".
This dialog comprises the settings for the power-on behavior of the instrument.
Power-On State
Selects the state which the RF output takes after the instrument is switched on.
"RF Off"
"Previous Setting"
Remote command:
:OUTPut[:STATe]:PON
User Manual 1176.8674.02 ─ 07
on page 327
on page 327
The output is deactivated when the instrument is switched on.
When the instrument is switched on, the output takes the same state
as it had when the instrument was switched off.
on page 328
Signal Generator Settings
RF Settings
on page 467
on page 467