8 General Instrument Settings and Instrument
Setup
This section describes the settings which do not directly affect signal generation.
Access:
► Select "SGMA-GUI > Instrument Name > Setup" and select the required dialog.

8.1 Internal Adjustments

The R&S SGT is accurate due to the integrated procedures for adjustments.
All internal adjustments for which no external measuring equipment is needed can be
started in the "Internal Adjustments" dialog. The adjustments with external measuring
equipment are described in the service manual.
During adjustment, a bar indicates the status of progress. If an error occurs, adjust-
ment is terminated and an error message is output in the info line.
Risk of invalid adjustment
In order to achieve correct adjustment of the instrument, make sure that the instrument
is warm before performing adjustments. The warm-up time is 30 minutes.
User Manual 1176.8674.02 ─ 07

General Instrument Settings and Instrument Setup

Internal Adjustments
243