Mode; I/Q Level Vp Emf; Couple I/Q Bias; Bias (Emf) - R&S SGT100A User Manual

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
"Single-Ended" ●
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:TYPE

Mode

Selects the mode for setting the outputs, independent or together.
"Fixed"
"Variable"
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:MODE

I/Q Level Vp EMF

Sets the output voltage for both signal components.
See
Table 7-4: Maximum I/Q level and overall output voltage
"I/Q Output
Type"
"Single ended"
"Differential"
Remote command:
[:SOURce<hw>]:IQ:OUTPut:LEVel

Couple I/Q Bias

In "Variable Mode", activates/deactivates coupling of bias setting of the I signal and the
Q signal component. With activated coupling, the settings of the I signal component
automatically apply to the Q signal component.
Remote command:
[:SOURce<hw>]:IQ:OUTPut[:ANALog]:BIAS:COUPling[:STATe]

Bias (EMF)

If a bias is enabled, a DC voltage is superimposed upon the I or Q signal.
User Manual 1176.8674.02 ─ 07
If "RF Envelope > On"
The envelope signal E is output at the I connectors.
If "RF Envelope > Off"
Single-ended output at the I/Q connectors.
You can define a bias between the output signal and ground.
The I/Q signal components settings are fixed to the value listed in
Table
7-3.
Table 7-3: I/Q Settings in "Mode = Fixed"
"I/Q Output Type"
"Single ended"
"Differential"
(R&S SGT-K16)
Enables modification of the I/Q signal components settings.
Chapter 7.3.2,
"Impairments",
Value range
"I/Q Level Vp (EMF)"
20 mV to 2 V
40 mV to 4 V
on page 427
"I/Q Level Vp
"Bias (EMF)"
(EMF)"
1 V
0 V
2 V
0 V
on page 428
on page 133.
Maximum overall output voltage V
"I/Q Level Vp (EMF)" + "Bias (EMF)" ≤ 4 V
0.5*"I/Q Level Vp (EMF)" + |"Bias (EMF)"| + 0.5*"Offset (EMF)"
≤ 4 V
on page 428
Signal Generator Settings
I/Q Analog Output Settings
"Offset (EMF)"
0 V
0 V
max
EMF
on page 427