8.14 Standby and Restart

See

8.15 Diagnostic and Tests

This section describes the settings provided for diagnostic and test purposes.

8.15.1 Self-test

1. To access this dialog, select "SGMA-GUI > Instrument Name > Diagnostic Tests >
Self-test".
2. To trigger a self-test, select "Self-test".
Performs a self-test on all installed hardware options.
The result of the self-test, succeeded or failed, is displayed. The list of the numeric
results of the performed test cases is protected by protection level 2.
SCPI command:
:TEST:ALL:STARt
:TEST:ALL:RESult?

8.15.2 Connection Test

The "Connection Test" dialog provides settings for testing the connection to an external
IQ device and performing a self-test of the baseband input.
► To access this dialog, select "SGMA-GUI > Instrument Name > Diagnostic Tests >
Connection Test".
User Manual 1176.8674.02 ─ 07
on page 311
Chapter 9.7, "How to Switch between the Operating
on page 488
on page 488
General Instrument Settings and Instrument Setup
Diagnostic and Tests
States",
on page 270.
263