:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type>
Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Also, it is recommended that you zero in regular intervals (at least once a day), if:
The temperature has varied more than about 5 °C.
The sensor has been replaced.
You want to measure very low power.
Note: The RF power source must be switched off or disconnected from the sensor
before zeroing.
Example:
Usage:
Manual operation:
:SLISt[:LIST]?
Returns a list of all detected sensors in a comma-separated string.
Return values:
<SensorList>
User Manual 1176.8674.02 ─ 07
SENSe, READ, INITiate and SLISt Subsystems
string
SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Query only
See
" Serial Number "
SENS1:ZERO
Executes zeroing.
Event
See
" Zero "
on page 226
String of comma-separated entries
Each entry contains information on the sensor type, serial num-
ber and interface.
The order of the entries does not correspond to the order the
sensors are displayed in the "NRP Sensor Mapping" dialog.