PCIe extension cables must fulfill the following requirements:
Single lane connectors
Max. cable length of 5 m.
For example: OSS-PCIe-CBL-x1 cable from One Stop Systems or 74576-000x cable
from Molex.
Connecting an external PC that does not support hot-plugging
1. Switch off the external PC and the instrument.
See also chapter "Switching the instrument On and Off" in the getting started man-
ual.
See also
2. Connect the instrument and the controller with the suitable cable.
3. Switch on the instrument.
4. Wait until the instrument has completed the booting (the "POWER" LED on the
instrument's front panel is constantly on).
5. Switch on the external PC.

10.3.5 Enabling Fast Settings

► To enable the special PCI express feature fast settings, enable the function
rssgt_UseFastSettings (ViSession instrumentHandle, ViBoolean
fastEnabled, ViBoolean asynchronousEnabled) included in the instru-
ment driver.
Settings for some parameters like level and frequency accelerate.
To disable the fast settings, call the function rssgt_UseFastSettings with argu-
ment fastEnabled=false.

10.4 Advanced Remote Control Using Fast Socket

Fast settings
The socket interface can be utilized not only to transfer text messages e.g. SCPI com-
mands but also to carry register based remote control messages. The latter mode is
called fast settings mode. The fast socket communication is based on the Ethernet pro-
tocol which does not support routing. Therefore a controller PC can only control devi-
ces within its own network segment using the fast socket method.
User Manual 1176.8674.02 ─ 07
Chapter 2.1.5, "Switching the Instrument On and
Network and Remote Control Operation
Advanced Remote Control Using Fast Socket
Off",
on page 21.