Sense<Ch>[:Power]:Filter:nsratio:mtime; Sense<Ch>[:Power]:Filter:sonce - R&S SGT100A User Manual

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Parameters:
<NSRatio>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:NSRatio:MTIMe <MTime>
Sets an upper limit for the settling time of the auto-averaging filter in the NSRatio
mode and thus limits the length of the filter. The filter type is set with command
SENSe<ch>[:POWer]:FILTer:TYPE.
Parameters:
<MTime>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:SONCe
Starts searching the optimum filter length for the current measurement conditions. You
can check the result with command :SENS1:POW:FILT:LENG:USER? in filter mode
USER (:SENSe<ch>[:POWer]:FILTer:TYPE).
Example:
Usage:
Manual operation:
User Manual 1176.8674.02 ─ 07
SENSe, READ, INITiate and SLISt Subsystems
float
Range:
0.001 to 1
Increment: 0.001
*RST:
0.01
SENS1:FILT:TYPE NSR
Selects fixed noise filter mode.
SENS1:FILT:NSR 0.2
Sets a noise content of 0.2.
See
"Noise Content"
on page 228
float
Range:
1 to 999.99
Increment: 0.01
*RST:
4
SENS1:FILT:TYPE NSR
Selects fixed noise filter mode.
SENS1:FILT:NSR .2
Sets a noise content of 0.2.
SENS1:FILT:NSR:MTIM 5
Limits the settling time to 5 seconds.
See
"Timeout"
on page 228
SENS1:FILT:TYPE USER
Selects user filter mode.
SENS1:FILT:SONC
Activates the search for the optimum filter length.
SENS1:FILT:LENG?
Returns the found optimum filter length.
Response: 128
Event
See
"Auto Once"
on page 228
Remote Control Commands
:
334

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents