R&S SGT100A User Manual page 398

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Parameters:
<Amplitude>
Example:
Manual operation:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:FREQuency <Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency>
Example:
Manual operation:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:OFFSet <Offset>
Sets the DC component.
Parameters:
<Offset>
Example:
Manual operation:
[:SOURce<hw>]:BB:ARBitrary:TSIGnal:RECTangle:SAMPles <Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
<Samples>
User Manual 1176.8674.02 ─ 07
float
Range:
0 to 1
Increment: 0.001
*RST:
0.800
Default unit: FS
BB:ARB:TSIG:RECT:AMPL 0.5
sets the amplitude of the test signal
See
"Amplitude"
on page 108
float
Range:
100 to depends on hardware
Increment: 0.01
*RST:
1000
SOURce:BB:ARBitrary:TSIGnal:RECTangle:FREQuency
100 kHz
sets the frequency of the test signal of 100 kHz.
See
"Frequency"
on page 107
float
Range:
-1 to 1
Increment: 0.001
*RST:
0
Default unit: FS
BB:ARB:TSIG:RECT:OFFS 0.5
sets the DC component
See
"Offset DC"
on page 108
integer
Range:
4 to 1000
*RST:
100
Remote Control Commands
SOURce Subsystem
367

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents