[:Source<Hw>]:Correction:cset:data:frequency:points; [:Source<Hw>]:Correction:cset:data:power; [:Source<Hw>]:Correction:cset:data:power:points - R&S SGT100A User Manual

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Parameters:
<Frequency>
Example:
Manual operation:
[:SOURce<hw>]:CORRection:CSET:DATA:FREQuency:POINts?
The command queries the number of frequency values in the selected table.
The numerical suffix at SOURce must not be used for this command.
Return values:
<Points>
Example:
Usage:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer <Power>
Transfers the level data to the table selected with
CSET[:SELect].
*RST does not affect data lists. The numerical suffix at SOURce must not be used for
this command.
Parameters:
<Power>
Example:
Manual operation:
[:SOURce<hw>]:CORRection:CSET:DATA:POWer:POINts?
Queries the number of level values in the selected table.
User Manual 1176.8674.02 ─ 07
Frequency#1[, Frequency#2, ...]
Range:
300 kHz to RFmax (depending on model)
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:FREQ 100MHz,102MHz,103MHz,...
enters the frequency value in the table ucor1.
See
" Edit User Cor. Data "
integer
Range:
0 to 10000
*RST:
0
CORR:CSET '/var/user/'
selects the table ucor1.
CORR:CSET:DATA:FREQ:POIN?
queries the number of frequency values in the table ucor1.
Response: 440
the table ucor1 contains 440 frequency values.
Query only
Power#1[, Power#2, ...]
CORR:CSET '/var/user/ucor1'
selects the table ucor1.
CORR:CSET:DATA:POW 1dB, 0.8dB, 0.75dB,...
enters the level values in the table ucor1.
See
" Edit User Cor. Data "
Remote Control Commands
SOURce Subsystem
on page 213
[:SOURce<hw>]:CORRection:
on page 213