Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:LENGth:AUTO?
Queries the current filter length in filter mode AUTO
FILTer:TYPE)
Return values:
<Auto>
Example:
Usage:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:LENGth[:USER] <User>
Selects the filter length for
a multiplier for the time window, a constant filter length results in a constant measure-
ment time. You can set values 1 and 2
The time window is fixed to 20 ms.
Parameters:
<User>
Example:
Manual operation:
:SENSe<ch>[:POWer]:FILTer:NSRatio <NSRatio>
Sets an upper limit for the relative noise content in fixed noise filter mode
SENSe<ch>[:POWer]:FILTer:TYPE). This value determines the proportion of intrin-
sic noise in the measurement results.
User Manual 1176.8674.02 ─ 07
SENSe, READ, INITiate and SLISt Subsystems
SENS1:POW:DISP:PERM:STAT ON
The permanent viewer is switched on.
See
" Permanent "
on page 226
float
Range:
1 to 65536
SENS1:FILT:TYPE AUTO
Selects auto filter.
SENS1:FILT:LENG:AUTO?
Queries the automatically set filter length.
Response: 1024
Query only
See
"Filter Length"
on page 227
SENS:POW:FILT:TYPE
n
.
float
Range:
1 to 65536
*RST:
1
SENS1:FILT:TYPE USER
Selects user filter mode.
SENS1:FILT:LENG 16
Sets a filter length of 16. The resulting measurement time is 640
ms (2x16x20 ms).
See
"Filter Length"
on page 227
Remote Control Commands
(:SENSe<ch>[:POWer]:
USER. As the filter length works as
(: