R&S SGT100A User Manual page 508

Sgma vector rf source
Table of Contents

Advertisement

®
R&S
SGT100A
Return values:
<All>
Example:
Usage:
:SYSTem:ERRor:CODE:ALL?
Queries all entries in the error queue and then deletes them. Only the error numbers
are returned and not the entire error text.
Return values:
<All>
Example:
Usage:
:SYSTem:ERRor:CODE[:NEXT]?
Queries the oldest entry in the error queue and then deletes it. Only the error number
is returned and not the entire error text.
Return values:
<Next>
User Manual 1176.8674.02 ─ 07
string
List of: Error/event_number,"Error/event_description>[;Device-
dependent info]"
If the queue is empty, the response is 0,"No error"
SYST:ERR:ALL?
queries all entries in the error queue.
Response: 0, 'no error'
No errors have occurred since the error queue was last read out.
Query only
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by
SCPI.
SYST:ERR:CODE:ALL
queries all entries in the error queue.
Response: 0
no errors have occurred since the error queue was last read out.
Query only
string
0
"No error", i.e. the error queue is empty
positive value
Positive error numbers denote device-specific errors
negative value
Negative error numbers denote error messages defined by
SCPI.
Remote Control Commands
SYSTem Subsystem
477

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents