R&S
SGT100A
9.5 How to Load and Play Waveform Files
Irrespective of the way a waveform file is generated, you can transfer it to the instru-
ment, load it in the ARB and play it.
To load and play a waveform file from the instrument
1. Transfer an externally created waveform file to the instrument, see
"Managing Files on the R&S
2. Select "Baseband > ARB > General".
3. Set "Waveform Location > Instrument".
4. Select "Load Waveform...". Navigate to the directory the file is stored in, select the
waveform file (e.g. /var/user/lte_waveform) and execute "Select".
The "ARB" dialog confirms that the waveform file is loaded.
5. Select "ARB > State > On".
The R&S SGT processes the waveform file.
To load and play a waveform file from an external computer
1. Connect the R&S SGT and a Windows PC with an installed R&S SGMA-GUI to the
same network.
2. On the remote PC, select "SGMA-GUI > Instrument Name > Baseband > ARB >
General".
3. Set "Waveform Location > Local file system ".
4. Select "Load Waveform...". Navigate to the directory the file is stored in, select the
waveform file (e.g. /var/user/lte_waveform) and execute "Select".
Note: You can also select an external USB connected to the PC at the available
local file directories.
The "ARB" dialog confirms that the waveform file is loaded.
5. Select "ARB > State > On".
The R&S SGT processes the waveform file.
9.6 How to Restore the LAN Connection to an Instrument
► If the LAN connection to an instrument configured to use a static IP address is lost,
press the LAN LED on the instrument's front panel for more than 3 seconds to
reset the LAN settings and to set the
User Manual 1176.8674.02 ─ 07
Performing Configuration Tasks
How to Restore the LAN Connection to an Instrument
SGT",
on page 46.
"Address Mode"
Chapter 5.2,
on page 257 to DHCP.
269