[:SOURce<hw>]:BBIN:ALEVel:EXECute
This command starts measuring the input signal. The measurement estimates the
crest factor and the peak power.
Example:
Usage:
Manual operation:
[:SOURce<hw>]:BBIN:CDEVice?
Digital Input only.
Indicates the ID of an externally connected R&S Instrument or R&S Device.
Return values:
<CDevice>
Example:
Usage:
Manual operation:
[:SOURce<hw>]:BBIN:CFACtor <CFactor>
This command enters the crest factor of the external baseband signal.
Parameters:
<CFactor>
Example:
Manual operation:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe <State>
Activates automatic adjustment of the baseband input signal.
The R&S SGT receives peak level, level and crest factor values directly from the con-
nected transmitter and recognizes changes automatically.
Parameters:
<State>
User Manual 1176.8674.02 ─ 07
BBIN:ALEV:EXEC
starts measuring the input signal.
Event
See
"Auto Level Set"
on page 240
string
BBIN:CDEV
queries the connected device ID.
Query only
See
"Connected Device"
float
Range:
0 to 30
Increment: 0.01
*RST:
0
Default unit: dB
BBIN:CFAC 10
enters a crest factor of 10 dB for the external baseband signal.
See
"Crest Factor"
on page 241
0 | 1 | OFF | ON
*RST:
0
Remote Control Commands
SOURce Subsystem
on page 238
401