"ON"
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
DAC Value
Enters a user-defined adjustment value for the internal reference frequency. This value
is not used unless "Adjustment Active > On" is selected. "0" represents the calibrated
state. The setting range depends on the reference oscillator type and its factory cali-
bration value.
Remote command:
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue

7.7.6 RF Level Settings

This section explains the level settings of the R&S SGT.
Configuring RF level
► To change the RF level of the selected instrument, perform one of the following:
a) Select "SGMA-GUI main panel > Level" and enter the desired value.
b) Select "SGMA-GUI > Instrument Name > RF > Level Settings > RF Level >
Changes of the RF level have an immediate effect on the output signal.
User Manual 1176.8674.02 ─ 07
A user-defined adjustment value is used. The value is entered under
DAC
Value.
This allows the frequency to be impaired freely, for example, to simu-
late a frequency error.
The instrument is no longer in the calibrated state. However, the cali-
bration value is not changed and the instrument resumes the calibra-
ted state after disabling the adjustment.
Level" and enter the desired value.
Signal Generator Settings
RF Settings
on page 471
on page 471
205