The X-axis-position adjusting knob shifts 0.5 mm per revolution, and the
Y-axis-position adjusting knob shifts 0.25 mm per revolution.
When the following result is obtained, calculate the calibration value from this
graph.
0.5
• Distance d [mm/V]
Enter the distance of 40.0 between the cantilever and the photodiode in this input
box.
• Reference V
Enter the Friction Force signal value that is obtained when the force applied to the
cantilever is zero in the Reference V0 [V] input box.
TMPM5200-2
The moving distance of the photodiode by turning the adjusting knob per
revolution is as follows:
X-axis (mm)
0.25
The distance is different depending on the model of the instrument:
[V]
0
FFM (V)
+10
v
1
x
2
x
1
v
2
—10
a
Y-axis (mm)
6 ANALYSIS OPERATION
X
x
x
=
1
2
V
V
1
2
6-115