Set a Registered Image
Sets the latest image or a specified logging image as a registered image.
Registered images are saved to the following directory on the SD card or PC Tool.
\Sensor name\REGIMAGE
The image specified as a registered image can be loaded as a measurement image with the Acquire
Registered Image command.
Acquire Registered Image p.401
A maximum of 1000 registered images (image registration number: 0 to 999) can be set.
*
For the PC Tool, the data is saved in the "\..\My Documents\OMRON FQ\" folder.
Command Format
MRC
SRC
(1 byte)
(1 byte)
28
0F
Response Format
MRC
SRC
(1 byte)
(1 byte)
28
0F
End Codes
End code (hex)
0000
FFFF
Parameter Descriptions
Registered image num-
ber
Registration source
Logging image number
400
Command Control
*
Sensor command code (4
bytes)
00108010
MRES
SRES
(1 byte)
(1 byte)
End code
Command execution ended normally.
Command execution ended in an error.
Specifies the image to be loaded as the measurement image.
The image is registered in advance with the Set a Registered Image command and assigned a number. (0
to 999)
Specifies the image to be set as the registered image.
0: Latest measurement image
1: Sensor logging image
This argument is only specified when setting the Sensor's logging images as registered images. (0 to 19)
If the registration source is 0: Latest measurement image, this argument is not required.
Registered image number
(4 bytes)
3 digits max.
Sensor command code (4 bytes)
00108010
Description
Registration
Logging image
source (4 bytes)
number (4 bytes)
1 digit max.
2 digits max.
FQ2-S/CH User's Manual