Measuring Open Calibration Data - Agilent Technologies E4991A Installation And Quick Start Manual

Impedance/material analyzer
Hide thumbs Also See for E4991A:
Table of Contents

Advertisement

Basic Operations for RF Devices Measurement
STEP 3. Calibration
Step 4. As shown in Figure 3-30, turn the 7-mm connector nut of the test head clockwise until the
7-mm connector sleeve is fully extended.
Figure 3-30
Extending 7-mm connector sleeve

Measuring OPEN Calibration Data

Use the 0 S (OPEN) standard to perform OPEN calibration by following these steps.
Step 1. As shown in Figure 3-31, turn the 0 S (OPEN) standard clockwise with the provided torque
wrench to connect it securely to the 7-mm terminal.
Figure 3-31
Connecting 0 S (OPEN) standard
Step 2. Click the
Figure 3-32
Step 2
Step 3. A √ mark appears on the left side of the
calibration data measurement.
62
button (Figure 3-32).
Meas Open
button upon completion of the OPEN
Meas Open
Chapter 3