33
Trigger Commands

:TRIGger:RUNT:POLarity

Command
:TRIGger:RUNT[{1 | 2}]:POLarity {POSitive | NEGative}
This command defines the polarity of the runt pulse as positive or
negative. The trigger source must be set using the :TRIGger:RUNT:SOURce
command.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:RUNT:POLarity?
The query returns the currently selected runt pulse polarity.
Returned Format
[:TRIGger:RUNT:POLarity] {POSitive | NEGative}<NL>