Removing And Installing Acquisiton Modules; Removing Modules - HBM Genesis GEN5i User Manual

Portable, integrated data acquisition system
Table of Contents

Advertisement

I2679-4.0 en
7.3

Removing and installing acquisiton modules

CAUTION
HBM uses state-of-the-art electronic components in its equipment. These
electronic components can be damaged by discharge of static electricity
(ESD). ESD damage is quite easy to induce, often hard to detect, and
always costly. Therefore we must emphasize on the importance of ESD
preventions when handling a GEN5i system, its connections or a plug-in
card.
CAUTION
The GEN5i Data Acquisition System is factory-calibrated as delivered to
the customer. Swapping, replacing or removing of boards may result in
minor deviations to the original calibration. The GEN5i system should be
tested and if necessary, calibrated, at one-year intervals or after any major
event that may affect calibration. When in doubt, consult your local
supplier.
7.3.1

Removing modules

CAUTION
Heatsink and other parts of the card/module may be hot when removed
just after switch-off.
To remove a module:
1
Shut down the GEN5i system and remove the power input cable.
2
Disconnect all cables from acquisition boards.